- 4, 8, or 16 channels
- 1 MHz aggregate 12 bit sampling. Resolution is 0.024% of full scale
- 13 recording durations, user selected: 132 msec to 1320 sec
- Digital Triggering from any channel, with +/- slope or window
- 6 full scale gain ranges, user selected: 0.4, 1, 2, 4, 10, and 20 volts full scale - with a 10mV/g accelerometer these equal 40g, 100g, 200g, 400g 1000g and 2000g full scale - higher or lower ranges by different sensor selection
- Streaming to HD mode saves multiple amplitude triggered records - for environment characterization and averaged or statistical SRSs
- Hardware calibration done once at factory, certificate provided - throughput calibrated to better than 1.5% of full scale
- Ultra high speed multiplexing - 1 µsec between channels
- Hardware signal conditioning - constant current at 18-24 VDC for voltage type sensors, DC inputs by BNCs
- LapCATIII interfaces to most notebook PCs through PCI Card Bus
- GHI can also supply compatible notebook PCs - Dell, IBM, Toshiba, and Compaq
- Furnished with WinCAT software - optionally SRS, Damage Boundary, and MilSPEC
- WinCAT Software may be combined with ProCAT in a single system
|
|
Click here to see a larger image.
Click here to download LapCAT III PDF.
|